Output / wedge factor determination
Output factor determination - comparison of the field factors calculated by TPS vs. measured
Select the ROI characteristics for the average dose reading (DFn = dose for the Field n) and the reference Field (Fr). The output factor is derived as: OF = DFn / DFr
You can generate an Output factor report from the results.
Wedge factor determination - comparison of the field factors calculated by TPS vs. measured
The wedge factor is derived as: Wedge Factor = DFn / DFr where DFn is the dose of Field n and DFr is the open field reference dose.
You can generate a Wedge factor report from the results.