Output factor report

Institute: DemoDemo Department: DemoDemoDemoDemo
Patient ID: OPEN__MACH ID QA: OPEN
Plan: OPEN___MACHINE LINAC: 6EX    6 MV
Detector: aS500 Acquisition date: 2008/05/09
Epiqa calib date : Epiqa calib depth : mix
Description:


Results

   X size [cm] Y size [cm] Gy/ 100MU ECL Gy/ 100MU PV Diff % OF ECL OF PV Diff %
20x20 20.00 20.00 1.225 1.221 -0.4 1.044 1.043 -0.1
20x6 6.00 20.00 1.170 1.177 0.6 0.996 1.005 0.9
10x10 10.00 10.00 1.174 1.171 -0.3 REF ECL REF PV





Notes:__________________________________________________________________
_______________________________________________________________________


Approvement:Physicist: