Which TPS and linac Epiqa supports? Article rated 3.7/5.0
At this moment Epiqa supports all Varian linacs and TPS Eclipse and Pinnacle (Philips).
Can I use Epiqa with Elekta or Siemens machine? Article rated 3.7/5.0
No, we have no plans to support those machines at this time.
Measurements in Epiqa shows poor conformity than other method of evaluation. Article rated 3.8/5.0
We suggest you to check your IC.dat file which you use for Epiqa evaluation if the data was measured/calculated correctly for your configuration. If you use mix configuration : Example for 6MV: So...
Using of demo mode Article rated 3.8/5.0
Epiqa always runs in demo mode, if there is no "license.dat" file in the "license" subfolder (default path: C:\Program Files\EPIdos\Epiqa\license). If you have created license an...
Sorting problem in evaluation of TrueBeam images Article rated 4.1/5.0
Linking of the images to fields using file names for TB has not been implemented for PV vs PV analysis yet (Athena module). We will insert this feature in one of new Epiqa versions.
Epiqa support of machine override (during image acquisition) possibility Article rated 3.7/5.0
In case that you decide to acquire images at other machine than machine in the plan, your result will be incorrect becasue EPiqa will read data to calculate dose from the "planned machine" r...
Log file in GB Article rated 4.1/5.0
By default the log files in txt format which are stored in log folder usually haven`t more than a couple of kB. If your log file is huge please check in Tools â?? Preferences â?? General 2 if the...
Is it possible to use Epiqa for other energies than X6 especially X18 and the electron energies (E6-E18)? Article rated 3.8/5.0
You can use Epiqa for all photon energies. You have to configure Epiqa for each energy, each dose rate (or the highest dose rate) and each setting of PV detector (d max, mix or build up mode) which yo...
field 10x10_50 Article rated 4.1/5.0
The 10x10 reference field is used to compensate for any offset due to daily variation in linac output and detector response. Automatic alignment is done on the acquired 10x10 field; in any case, smal...
Gaussian convolution - how is it used Article rated 2.3/5.0
This allows you to perform a Gaussian convolution on the PV measured data. It can be effective to see the effect of the sampling of the data. In general, comparing Eclipse matrix (where you have a ce...
Epiqa in FFF mode Article rated 4.1/5.0
TrueBeam currently doesnâ??t support portal imaging of FFF beams. Therefore it is not possible to analyse those beams in Epiqa either. But you can use Epiqa for flattened TrueBeam energies.