Measurements in Epiqa shows poor conformity than other method of evaluation.
Solution
We suggest you to check your IC.dat file which you use for Epiqa evaluation if the data was measured/calculated correctly for your configuration. If you use mix configuration :
Example for 6MV:
Source Detector Distance; SDD = 100 cm, this corresponds to the detector vertical position reading 0.0 cm.
Source Surface Distance; SSD = SDD â?? dmax = 98,5 cm, this a correct SSD set in Eclipse for this example.
dmax (in water) = 1.5 cm.
d(Portal Vision) = 1.2 cm, this is the real depth of the aSi layer inside of Portal Vision detector (refer to manual on pages 22-24 and 40-41)
Do you use your own measured beam data for Eclipse calculation? Sometimes customers use Golden beam data from Varian and this can results in a discrepancies in Epiqa evaluation, because standard 2D matrices doesn`t have so high resolution as EPID and they find this problem only with Epiqa.