Solution The 10x10 reference field is used to compensate for any offset due to daily variation in linac output and detector response.
Automatic alignment is done on the acquired 10x10 field; in any case, small adjustment through a minimization process is added: this allows to take into account small imager shifts during the RapidArc plan delivery, e.g., for the average displacement of the arm during the rotation.
The 10x10 is used to find the radiant center of the image; then for each single field acquired few pixel of additional shift could be introduced according to the results of the minimization process.
In principle it would not need a specific gantry geometry for 10x10 acquisition, and it is very common acquired it at gantry 0, as first field. You can check in any case, if the position of PV is still good in terms of arm performance after rotation of the gantry (e.g. 0,0,0); if it is not the case (e.g. with the R-arm, while with exact arm should not), please re-position PV (e.g.0,0,0).
You can check also in the properties of your images the actual registered arm positions.
 