Gaussian convolution - how is it used
Solution
This allows you to perform a Gaussian convolution on the PV measured data.
It can be effective to see the effect of the sampling of the data. In general, comparing Eclipse matrix (where you have a certain dose calculation grid) and the measurement where the resolution is higher, you can see differences due to the low dose sampling of calculation. Usually the agreement between a convoluted measurement and dose calculation matrices is higher - after a convolution, the measured dose distribution is smoother. Clearly, this tool has to be used carefully, thinking that you are delivering peaks and valleys that are not fully calculated. We would not suggest to use it to â??improveâ?� GAI results but rather to observe sensitivity of the results to detector resolution vs. calculation matrix resolution.