Epiqa support of machine override (during image acquisition) possibility

Solution In case that you decide to acquire images at other machine than machine in the plan, your result will be incorrect becasue EPiqa will read data to calculate dose from the "planned machine" rather than "machine of acquisition". THis is due to the fact, that Epiqa select dat based on the infomration in the plan.

To implement support for such scenario is rather complicated. There are few complications with this approach such as different imager models (IDU11 and IDU20, aS-500 vs. aS1000). Further we see that the response of the imager from different sites almost always slightly differs most probably because of small changes in energy. There is also the problem that image DICOM header doesn't contain any information about the machine at which it was acquired. So automatic solution will most probably not be possible.
 
Was this article helpful? yes / no