RapidArc patient QA
Use your portal imager as absolute dosimeter to verify RapidArc plans following these steps:
- Replace patient 3D model with water phantom and reset gantry angles to zero for dose calculation
- Calculate verification plan using YOUR CLINICAL dose calculation algorithm
- DICOM export dose plane at dmax as a reference matrix
- Deliver the RapidArc plan normal way while acquiring integrated image
- Compare calculated versus acquired dose matrices using Epiqa
Benefits of using Epiqa for RapidArc plan QA:
- Simple QA plan preparation
- Instant detector preparation for measurement – just deploy Portal Vision arm
- Easy QA plan delivery – can be performed by therapist
- Fast data export – export only 4 DICOM objects
- Quick results evaluation. In just few minutes user can print final protocol for approval
- The method verifies not just the delivery device but also TPS performance
White paper: Hrbacek J.: Verification of RapidArc® plan using EpiqaTM
Last update on 06. January 2011